Standardization of CalyculinA induced PCC assay and its advantages over Okadaic acid PCC assay in Biodosimetry applications
Rajesha K NairyNarayana YerolNagesh N BhatUtkarsha DesaiKapil ShirsathUsha YadavRajesh K. ChaurasiaSapra B K
著者情報
ジャーナル フリー HTML

2016 年 58 巻 6 号 p. 563-569

詳細
抄録

Objective: In the present study an attempt was made to estimate coefficients of dose response curves for PCC aberrations induced by CalyculinA and Okadaic acid, using 60Co gamma radiation and 8 MeV pulsed electron beam for biodosimetry application. Materials and Methods: The modified method outlined by Puig et al. 2013 was used to conduct Calyculin A and Okadaic acid induced PCC assay in human blood lymphocytes.Chemical treatment was given for the last 1 h of a 48 h culture. The study was carried out in the dose range 2.5 to 20 Gy using 60Co gamma rays and 8 MeV pulsed electron beam. Results and Conclusions: Results show a linear dose dependent increase with a slope of 0.047 ± 0.001 from Calycalin A PCC and 0.048 ± 0.002 form Okadaic acid PCC. The slope of the fragments curve was 0.327 ± 0.006 from Calyculin A and 0.328 ± 0.006 from Okadaic acid PCC. Further, dose calibration studies were carried out for 8 MeV electron using Calyculin A PCC assay and the obtained slope from ring yield was 0.054 ± 0.002 and 0.427 ± 0.009 from fragment yield.

著者関連情報

この記事は最新の被引用情報を取得できません。

2016 by the Japan Society for Occupational Health
前の記事 次の記事
feedback
Top