2024 Volume 66 Issue 1 Pages 70-74
Purpose: To clarify the magnification error caused by the degree of tilt of the incisor and the elevation of the X-ray focus position, and the verification effect of magnification correction when performing vertical dual-exposure panoramic radiography.
Methods: Panoramic radiographic images of a phantom embedding 26 steel balls were taken at different heights (0, 5, 10, 15, and 20 mm) and tilt angles (0°, 10°, 20°, and 30°) to evaluate vertical magnification in each condition. Error and correlation coefficients in the vertical magnifications were calculated between the measured and theoretical magnification values.
Results: The more the steel ball phantom was tilted, the more the images of the uppermost steel balls were laterally stretched. In the vertical direction, image magnification also influenced the tilt angle of the object in the incisal region. The range of error in vertical magnification was −0.35-0.30%. The Spearman’s rank correlation coefficient between the measured and theoretical magnification value was 0.983.
Conclusion: Vertical magnification correction has the potential to improve image quality when merging panoramic radiographs in vertical dual-exposure panoramic radiography.
Panoramic radiography (PR) is an imaging method that scans the dental arch using slit imaging and tomography [1,2]. When imaging incisors with this method, X-rays are passed through the head from behind. The X-rays pass through the cervical vertebrae and the incisors are projected onto the detector. A tomographic layer is set on the incisor, establishing a clear image. However, because the cervical vertebrae deviate greatly from the tomographic layer, they are superimposed on the incisor as a ghost image [3,4,5].
The ghost images of the cervical vertebrae and intervertebral space (IVS) alternate between radio-opaque and radiolucent images in the incisal region. When a radiolucent image overlaps with the apex of the incisor, it is difficult to distinguish a radiolucent ghost image from a periapical lesion [6]. In these cases, intraoral radiographs are generally taken. However, during the Covid-19 pandemic, intraoral radiography was discouraged to prevent the spread of infection [7,8,9,10].
To solve the problem of ghost images of the IVS overlapping the incisors, Kato et al. [11] proposed the vertical dual-exposure PR method. In this method, the first PR is taken in a conventional position, and the second is taken with the X-ray focus raised by 5 to 20 mm. Raising the X-ray focus moves the position of the ghost image of the IVS downward relative to the incisors. Although the ghost image of the IVS overlaps the incisors, the position of the ghost image shifts between the first and second PR images. The two PR images taken with the X-ray focus at different heights are then merged by the least squares method. Kato et al. [11] concluded that the vertical dual-exposure PR method can reduce the negative effects of ghost images of the cervical vertebrae and IVS.
However, when the X-ray focus is raised, the image of the incisors is vertically distorted because of the change in the angle of the incident X-ray beam to the incisors compared with the conventional position. Vertical dual-exposure PR therefore provides a poor merged image when there is a large difference in the length of the incisors between the two PR images [11]. Alternatively, the length of the incisors on PR depends on the tilt of the incisors in the labio-palatal direction because the magnification of the incisor varies as the distance between the incisal edge and the X-ray focus differs from the distance between the apex of the incisor and the X-ray focus in both conventional PR and vertical dual-exposure PR. These effects may cause distortion of the incisors in the vertical direction when the images of the incisors are merged in vertical dual-exposure PR. However, this effect has not been elucidated.
The aims of this study were 1) to clarify the magnification error caused by the tilt of the incisor and the elevation of the X-ray focus position, and 2) to assess the verification effect of magnification correction in the vertical direction when performing vertical dual-exposure PR.
Magnification of incisors depends on the horizontal positional relationship between the X-ray focus, the object, and the detector as shown in Fig. 1. When the X-ray focus and detector are raised and/or the incisors are tilted, the magnification value also varies. When raising the X-ray focus and detector from the original position (0 mm: F0) to H mm (FH), object length (OL: distance between a-point [Oa] and b-point [Ob] of the object in Fig. 1) on the detector changes from IF0 to IFH, indicating that the length of the incisor on the detector is slightly shortened. IF0 and IFH were calculated with the following formula (see Fig. 1):
IF0 = OLcosθ × dF-De/dF-Ob
IFH = OLcosθ × dF-De/dF-Ob − H × (dOa-De/dF-Oa − dOb-De/dF-Ob)
θ, tilt angle; dF-De, horizontal distance between X-ray focus and detector; dF-Oa, horizontal distance between X-ray focus and Oa; dOa-De, horizontal distance between Oa and detector; dF-Ob, horizontal distance between X-ray focus and Ob; dOb-De, horizontal distance between Ob of the object and detector.
The actual values of dF-De, dF-Oa, dOa-De, and OL were 518.0 mm, 398.5 mm, 119.5 mm, and 30.0 mm in this study, respectively. Other items were variable according to setting of H and θ. The theoretical vertical magnification of the object was 1.30 times at Oa. Finally, the theoretical magnification value (TMV) between the PR images taken at two different heights (F0 and FH) was calculated with the following formula:
TMV = IFH/IF0
= 1 – H/dF-Oa × tanθ
Both Oa and Ob are projected below the detector by raising the X-ray focus. Tilting makes Ob closer to the focus than Oa, and the magnification of Ob becomes greater than that of Oa. Therefore, Ob moves downward more than Oa. As a result, the size of the object on the detector is shortened by raising the focus. The theoretical magnification value was calculated from the geometrical relationship between X-ray focus, object and detector, and tilt angle and height of X-ray focus. F0, X-ray focus and detector at original position; FH, X-ray focus and detector at position raised by H mm; H, height of X-ray focus and detector; IF0, image length taken at F0; IFH, image length taken at FH; Oa, a-point of the object (incisal edge); Ob, b-point of the object (apex), OL, object length (distance between Oa and Ob); dF-De, horizontal distance between F and detector; dF-Oa, horizontal distance between X-ray focus and Oa; dOa-De, horizontal distance between Oa and detector; dF-Ob, horizontal distance between X-ray focus and Ob; dOb-De, horizontal distance between Ob and De; θ, tilt angle of the object
The steel ball phantom, which was originally made and has been used for maintenance of the PR apparatus, shown in Fig. 2A was used as the object. Twenty-six steel balls with a diameter of 0.5 mm were embedded in an acrylic plate and were arranged at equal intervals along a 50-mm-long straight line. The position of 30 mm from the top of the uppermost steel ball was set as the center of rotation when tilting.
Human head phantomA human head phantom (SE-2, Osaka Kasei Co., Osaka, Japan) with cervical vertebrae was used as the object.
A: overview of the steel ball phantom. Steel balls 0.5 mm in diameter were embedded in a 50-mm-long acrylic plate at intervals of 2 mm. The center position was placed 30 mm from the top (Oa). The uppermost steel ball was designated as Ob. B: the image merged from photographs taken from 0°to 10°, 20°, and 30° with the Oa as the center.
Veraviewepocs X550 (J. Morita Co.) was used for imaging. The exposure conditions were 60 kV, 1 mA and 15 s, with an additional filter of 0.2 mm of copper plate for the steel ball phantom. The center position was 30 mm from the top of the uppermost steel ball. This center position was aligned to the tomographic layer with the lateral, central, and horizontal laser beam. It was then tilted from 0° to 10°, 20°, and 30° toward the X-ray focus side (Fig. 2B). PRs were also taken with the height of the X-ray focus raised from 0 mm to 5, 10, 15 and 20 mm.
For the human head phantom, the Frankfurt horizontal plane of the human head phantom was set parallel to the floor after midline positioning. The lateral laser beam for anteroposterior positioning of the tomographic layer was fixed at the left maxillary canine of the human head phantom. The exposure conditions were 80 kV, 5 mA and 15 s– half the conventional clinical conditions in milliampere-seconds– with no additional filter. Two PRs were taken with the X-ray focus at heights of 0 mm and 20 mm.
Image processing to construct subtracted and merged imagesOriginal software developed by C# (Microsoft, Redmond, WA, USA) was used to subtract and merge images by applying the least squares method [12,13,14,15,16,17]. The PR images taken at H0 were matched to those taken at heights of 5, 10, 15, and 20 mm by tilting from 0° to 10°, 20°, and 30°. At these positions, subtracted and merged images were constructed and exported as 8-bit grayscale bitmaps.
Calculation of magnification value and vertical magnification correctionUsing the original software and applying the least squares method, the magnification of the vertical direction of the PR image compared with the original PR image taken at H0 at each tilt angle was calculated as the measured magnification value (MMV) at each height position and tilt angle. The difference between the TMV and MMV was calculated as the error.
The subtracted and merged images were created with the original PR image taken at H0 and the PR image after magnification correction in the vertical direction with the MMV taken at each height of the X-ray focus. This process is referred to as the vertical magnification correction.
For the human head phantom, subtracted and merged PR images were obtained from the two PRs taken at X-ray focus heights of 0 mm and 20 mm. Vertical magnification correction was performed with the MMV measured as above.
Statistical analysisSpearman’s rank correlation coefficient was calculated between TMV and MMV using SPSS (IBM Corp., Armonk, NY, USA). A P value of less than 0.05 was considered to indicate a statistically significant difference.
Figure 3 shows images taken with the X-ray focus at the conventional height of 0 mm and the steel ball phantom tilted at 0°, 10°, 20°, and 30°. As the phantom was tilted, it expanded laterally as the upper steel ball deviated from the tomographic layer.
Figure 4 shows an image with the X-ray focus raised by 5, 10, 15, and 20 mm at each angle of 0°, 10°, 20°, and 30°. Figure 5 shows the subtracted PR images obtained by matching the positions using the least squares method. As the tilt angle increased, the X-ray focus position also increased, and as the position of the steel ball increased, the deviation of images shown as black and white horizontal lines also increased.
Table 1 shows the theoretical and actual correction of the magnification ratio at each tilt angle and height of the X-ray focus. The maximum error was 0.30%, and the minimum was −0.35%. Figure 6 shows the regression line of the TMV and MMV obtained by the least squares method at each tilt angle and height of the X-ray focus in the steel ball phantom experiment. The Spearman’s rank correlation coefficient was 0.983 (P < 0.05). Figure 7 shows the subtracted images between images at F0 and images with vertical magnification correction at FH. The matching area increased from the uncorrected image shown in Fig. 5. The merged images before and after vertical magnification correction also substantially match the merged images before vertical magnification correction (Fig. 8).
PR images of the human head phantom before and after vertical magnification correction are shown in Fig. 9. A white line was observed on the incisal edge of the maxillary incisor in the subtracted PR image (Fig. 9C) before correction, but it was alleviated after correction of the magnification ratio in the vertical direction (Fig. 9D).
The more the steel ball phantom was tilted, the more the images of the upper steel balls were laterally stretched and blurred.
The more the steel ball phantom was tilted, the more the images of the upper steel balls were laterally stretched and blurred.
The greater the tilt angle, the higher was the X-ray focus position, and the higher the position of the steel ball, the greater was the vertical displacement of the subtracted image.
F0, X-ray focus and detector at original position (0 mm); FH, X-ray focus and detector at a height of H mm from F0
Tilt angle | 0° | 10° | 20° | 30° | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Height of X-ray focus | TMV | MMV | error (%) | TMV | MMV | error (%) | TMV | MMV | error (%) | TMV | MMV | error (%) |
5 mm | 1.0000 | 0.9980 | 0.20 | 0.9978 | 0.9980 | −0.02 | 0.9954 | 0.9970 | −0.16 | 0.9928 | 0.9940 | −0.12 |
10 mm | 1.0000 | 0.9980 | 0.20 | 0.9956 | 0.9950 | 0.06 | 0.9909 | 0.9930 | −0.21 | 0.9855 | 0.9890 | −0.35 |
15 mm | 1.0000 | 0.9970 | 0.30 | 0.9934 | 0.9950 | −0.16 | 0.9863 | 0.9880 | −0.17 | 0.9783 | 0.9790 | −0.07 |
20 mm | 1.0000 | 0.9990 | 0.10 | 0.9912 | 0.9930 | −0.18 | 0.9817 | 0.9830 | −0.13 | 0.9710 | 0.9730 | −0.20 |
TMV, theoretical magnification value; MMV, measured magnification value. The error is a subtraction of the MMV from the TMV.
The Spearman’s rank correlation coefficient was 0.983, showing excellent correlation.
F0, X-ray focus and detector at original position (0 mm); FH, X-ray focus and detector at a height of H mm from F0
Blurring in the vertical direction is significantly improved after vertical magnification correction
M, merged image; M+VMC, merged image with after magnification correction
A: cropped panoramic radiographs of the human head phantom taken at a height of 0 mm. B: cropped panoramic radiographs of the human head phantom taken at a height of 20 mm. C: subtracted image between Figs. 9A and B. D: subtracted image between Figs. 9A and B with vertical magnification correction. E: merged images of Figs. 9A and B. F: merged images of Figs. 9A and B with vertical magnification correction, The horizontal white lines on the incisal edge of the maxillary incisors in C (arrow) disappeared after vertical magnification correction as shown in D. The incisal edge of the mandibular canine became visible in D (arrowhead) after vertical magnification correction.
Diagnosis in the incisal region has been hindered by overlapping ghost images of cervical vertebrae and the IVS on PR images [4]. In particular, when the ghost image of the IVS overlapped the apex of the incisors, periapical lesions were sometimes difficult to diagnose. Kato et al. [11] proposed vertical dual-exposure PR as a method to reduce ghost imaging of the cervical vertebrae and IVS. In this method, PR images are taken twice with the X-ray focus at different heights, and these images are merged. Raising the position of the X-ray focus makes the ghost image of the IVS shift downward to the apex of the incisor. The basic principle is the same as the eccentric projection method in intraoral radiography.
A merged PR image is obtained using the least squares method so that the images of the incisor on the two PRs match. The least squares method shifts the position of the image until the minimum value of the sum of the squares of each subtracted pixel value is reached [12,13,14,15,16,17], and is used in the energy subtraction method and dual imaging plate (DIP) method. Sekiguchi et al. [17] reported that DIP intraoral radiography can reduce noise and artifacts such as scratches and dust. In DIP intraoral radiography, the front and back imaging plate images were merged. Theoretically, the geometrical positional relationship between the front and back imaging plate images is substantially the same. In fact, the imaging plate is scanned with a laser beam while moving in a longitudinal direction. It became clear that the geometric positional relationship between the first front imaging plate and the second back imaging plate images did not match because the speed of movement of the imaging plate in the long axis direction changes because of slippage. The positional relationship between the two images was improved by correcting the magnification ratio in the longitudinal direction.
The PR used in this study has a semiconductor charged coupled device as a detector, and therefore it is assumed that there are no geometrical strain changes caused by the positional relationship of the detector [18,19]. PR has the two characteristics of slit imaging and tomography. When the object is in the tomographic layer, a sharp image can be produced where the tomographic layer coincides with the object. Alternatively, lateral direction distortion and blurring of the object images may occur outside of the tomographic layer. Therefore, the images of the steel balls that were closer to the X-ray focus than the tomographic layer are blurred and stretched in the lateral direction. For this reason, as shown by the arrow of Ob in Fig. 3, the larger the tilt angle and the higher the position of the steel ball, the more the image is stretched in the lateral direction. In contrast, when the object is outside of the tomographic layer, the projected image will be narrow in the lateral direction with blurring. Thus, the image of the steel ball located at the bottom in Fig. 3 became narrower in the lateral direction as the tilt angle increased.
There is no tomographic effect in the vertical direction in PR. The magnification ratio in the vertical direction is determined by the positional relationship between the detector, the object, and the X-ray focus, as in general radiography. Therefore, as the object becomes closer to the detector, the magnification decreases. Additionally, as the object becomes closer to the X-ray focus, the magnification increases. Therefore, as shown in Fig. 1, when the object is tilted, Ob is closer to the X-ray focus than Oa, so the magnification of Ob is greater than that of Oa. Furthermore, when the X-ray focus is raised, as shown in Fig. 1, the closer the Ob to the focus, the greater is the shifting distance than Oa. For this reason, if the X-ray focus is raised by a distance of H and the object is tilted as shown in Fig. 1, the height of the object on the detector in the vertical direction is shorter.
A PR image of an actual steel ball phantom is shown in Fig. 4. It is unclear from the image whether the vertical length is shortened. The subtracted images of the steel balls between the image at the conventional height of 0 mm and the images at heights of 5, 10, 15, and 20 mm in Fig. 5 demonstrate that the larger the tilt angle, the higher the elevation of the X-ray focus, and the closer to the upper edge of Ob from Oa as the center, the larger was the gap. The resulting gap is displayed as horizontal lines of black and white. The merged image without vertical magnification correction in Fig. 8 demonstrates that the larger the tilt angle, the higher the elevation of the X-ray focus, and the closer the proximity to the upper edge of the steel ball, the more blurred was the image in the vertical direction. After the correction, the black and white lateral lines in the vertical direction are smaller than those in Fig. 5 before correction. Blurring in the vertical direction was also reduced in the merged image (Fig. 8).
In the subtracted image of the human head phantom, the position of the incisal edge of the maxillary incisor did not accurately overlap the conventional position at X-ray focus heights of 0 mm and 20 mm, and thus it was observed as a white line (Fig. 9C). This was alleviated by the correction (Fig. 9D). It seems that the vertical height caused by the tilt of the incisor has been corrected. However, in the corrected differential image, the incisal edge of the mandibular canine was observed as a white line (Fig. 9D). This may be because the incisor of the mandible is located more lingually than the incisor of the maxilla, and the degree of magnification differs from that of the maxilla. As described above, these findings establish that the images are less likely to overlap because the magnification ratio in the vertical direction differs as the tilt of the incisor and the elevation of the X-ray focus increase. Additionally, it was clarified that the method of correcting the magnification ratio in the vertical direction by the least squares method agrees well with the TMV, and that the corrected images can create significantly more accurate merged images.
However, by tilting the steel ball phantom of the object, the object was positioned outside the tomographic layer. As a result, blurring occurred. This blur did not change with this correction. It was thought that this blurring in the lateral direction could be improved by using tomosynthesis [20,21]. Because a charged coupled device was used for the sensor in this experiment, it was not possible to change the tomographic layer after taking the PR. In PR equipped with a complementary metal-oxide semiconductor sensor, it would be possible to convert the tomographic layer by synthesizing it even after taking the PR image [21]. Future experiments should be conducted with PR with synthesizing capability.
In conclusion, this study showed that the tilt angle of the object in the incisal region was correlated with the amount of laterally stretched blur by changing the position of the tomographic layer and the object. In the vertical direction, image magnification was also influenced by the tilt angle of the object in the incisal region. Vertical magnification correction has the potential to improve image quality when merging panoramic radiographs in vertical dual-exposure PR, as well as to reduce the ghost imaging of the cervical vertebrae and intervertebral space. When it is necessary to obtain details of the incisors and their peripheral anatomical structures, the vertical dual-exposure PR method with vertical magnification correction could be a good choice for patients with gag reflex and severe trismus, and for pediatric patients who cannot undergo intraoral radiography, as well as under pandemics such as Covid-19.
The authors have no conflicts of interest to declare.
This work was supported by JSPS KAKENHI Grant Number JP22K10133.
The authors thank Helen Jeays, BDSc AE, from Edanz (https://jp.edanz.com/ac) for editing a draft of this manuscript.