Fracture load of zirconia implant supported CAD/CAM resin crowns and mechanical properties of restorative material and cement
Nadja Rohr Sabrina MärtinJens Fischer
Author information
JOURNAL OPEN ACCESS

2021 Volume 65 Issue 4 Pages 502-508

Details
Abstract

Purpose To test if resin CAD/CAM materials should be considered for zirconia implants and how their mechanical properties affect the fracture load.

Methods Fracture load of molar crowns of CAD/CAM materials (VITA CAD-Temp [CT], Cerasmart [CS], Lava Ultimate [LU], Pekkton Ivory [PK]) on zirconia implants (ceramic.implant, 4.0 mm) fixed either with no cement, temporary cement (Harvard Implant semi-permanent [HIS]), self-adhesive (VITA Adiva S-Cem [VAS]) or either one of two adhesive cements (Multilink Automix [MLA], VITA Adiva F-Cem [VAF]) was analyzed. The restorative materials were characterized by their flexural strength, fracture toughness, elemental composition and organic/inorganic ratio while compressive strength of the cements was measured.

Results For the fracture load significantly highest mean values were found overall for PK (2921 ±300 N) > LU (2017 ±499 N) > CS (1463 ±367 N) = CT (1451 ±327 N) (p > 0.05). When analyzing the effect of the cement on the fracture load the overall ranking was VAF (2245 ±650 N) ≥ MLA (2188 ±708 N) ≥ VAS (2017 ±563 N) > HIS (1757 ±668 N) = no cement (1595 ±757 N) (p <0.05), meaning fracture load increased with the compressive strength of the cements. Additionally, a linear trend was found between the fracture load and the fracture toughness of the restorative materials.

Conclusions All restorative materials exhibited fracture load values similar or higher than lithium disilicate tested previously. Fracture load of CT, CS and LU can be significantly increased when an adhesive cement with a high compressive strength is used.

Content from these authors
© 2021 Japan Prosthodontic Society

This is an open-access article distributed under the terms of Creative Commons Attribution License 4.0 (CCBY 4.0), which allows users to distribute and copy the material in any format so long as attribution is given to the Japan Prosthodontic Society.
https://creativecommons.org/licenses/by/4.0/