Masseter and digastric muscle activity evaluation using a novel electromyogram that utilizes elastic sheet electrodes
Akio FusayamaTomoaki MamenoMasahiro Wada Kazuhiro MurakamiToshikazu NezuShinya TokuonoShusuke YoshimotoTakafumi UemuraTsuyoshi SekitaniKazunori Ikebe
Author information
JOURNAL OPEN ACCESS
Supplementary material

2023 Volume 68 Issue 1 Pages 122-131

Details
Abstract

Purpose: To evaluate the reproducibility and reliability of a novel electromyogram (EMG) device with a flexible sheet sensor for measuring muscle activity related to mastication and swallowing.

Methods: We developed a new EMG device made of elastic sheet electrodes to measure the masseter and digastric muscle activities for evaluating mastication and swallowing. To examine the measurement reproducibility of the new EMG device, masseter muscle activity was analyzed using the intraclass correlation coefficient (ICC). Further, we measured the maximum amplitude, duration, integrated value, and signal-to-noise ratio (SNR) using the new EMG device and conventional EMG devices and evaluated the reliability using ICC and Bland−Altman analysis.

Results: We confirmed high ICC (1,1) and ICC (2,1) scores (0.92 and 0.88, respectively) while measuring the reproducibility of the new EMG device. When compared to the active electrode EMG device, we found a high correlation for the maximum amplitude (0.90), duration (0.99), integrated values (0.90), and SNR (0.75), with no observation of significant fixed errors. Moreover, the regression coefficient was not significant for any of the evaluation items and no proportional error was observed. Compared with the passive electrode EMG device, the maximum amplitude and duration were highly correlated (0.73 and 0.89). In addition, the SNR exhibited a significant fixed error. In contrast, the regression coefficient was not significant for any of the evaluation items and no proportional error was observed.

Conclusions: Our results suggest that the new EMG device can be used to reliably and reproducibly evaluate muscle activity during mastication and swallowing.

Content from these authors
© 2024 Japan Prosthodontic Society

This is an open-access article distributed under the terms of Creative Commons Attribution-NonCommercial License 4.0 (CC BYNC 4.0), which allows users to distribute and copy the material in any format as long as credit is given to the Japan Prosthodontic Society. It should be noted however, that the material cannot be used for commercial purposes.