Reliability of isometric knee extension muscle strength measurements made by a hand-held dynamometer and a belt: a comparison of two types of device
Munenori Katoh
Author information
JOURNAL FREE ACCESS

2015 Volume 27 Issue 3 Pages 851-854

Details
Abstract
[Purpose] The purpose of the present study was to compare the reliability of 2 hand-held dynamometers (HHD-1, 2) with different designs, by performing isometric knee muscle extension measurements two times each. [Subjects] The subjects were 40 young healthy adults. [Methods] The reliability of the measurements was examined using Bland-Altman analysis. [Results] Bland-Altman analysis found a fixed bias in measurements made by HHD-1 with an average limits of agreement (LOA) value of −2.1 kgf. For HHD-2, only random errors were detected, and the minimal detectable change (MDC) was 11.4 kgf. Fixed biases were observed between the two devices with an average LOA value of 2.2 kgf. When the bodyweight ratio was used, fixed biases were observed in measurements made by both devices, and the average value of LOA was −0.03 kgf/kg. The comparison of the two devices revealed only random errors, and MDC was 0.22 kgf/kg. [Conclusion] For HHD measurements using these two devices, the appropriate number of measurements is two times, and comparison of measurement values between the two devices should be avoided.
Content from these authors
© 2015 by the Society of Physical Therapy Science. Published by IPEC Inc.

This article is distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (by-nc-nd) License.
https://creativecommons.org/licenses/by-nc-nd/4.0/
Previous article Next article
feedback
Top