Journal of Robotics and Mechatronics
Online ISSN : 1883-8049
Print ISSN : 0915-3942
ISSN-L : 0915-3942
Special Issue on MEMS for Robotics and Mechatronics
Effect of Inner Diameter and Anticoagulation Coating in a Microneedle on its Blood Suction Performance
Seiji AoyagiRyosuke NomuraTomokazu TakahashiMasato Suzuki
著者情報
ジャーナル オープンアクセス

2020 年 32 巻 2 号 p. 362-370

詳細
抄録

As a part of the development of a minimally invasive hollow microneedle designed to mimic a mosquito proboscis, we evaluated the relationship between the needle inner diameter (ID) and blood sucking performance. If the ID is thinned to reduce pain upon piercing skin, blood could clog the tube owing to coagulation, and a sufficient volume of blood might not be obtained. In this study, laser stereo-lithography is used to easily fabricate microtubes of several sizes, at 20–50 μm ID and a fixed length of 500 μm, through which human whole blood is sucked by a vacuum pump. The results indicate that the ID of the tube must be at least 20 μm to prevent hemolysis and at least 50 μm to enable extraction of 200 μL of blood, which is necessary for general blood tests. Moreover, anticoagulant coating applied on the inner wall prevents the clogging of blood and increases the volume of extracted blood.

著者関連情報

この記事は最新の被引用情報を取得できません。

© 2020 Fuji Technology Press Ltd.

This article is licensed under a Creative Commons [Attribution-NoDerivatives 4.0 International] license (https://creativecommons.org/licenses/by-nd/4.0/).
The journal is fully Open Access under Creative Commons licenses and all articles are free to access at JRM Official Site.
https://www.fujipress.jp/jrm/rb-about/
前の記事 次の記事
feedback
Top