2007 Volume 14 Issue 1 Pages 59-68
The high voltage generation, measurements, and calibration were described. Measurements of 10 ppm in uncertainty for an AES was attained. This was achieved by compensating a temperature coefficient and accounting a power coefficient of the resistors. A standard high-voltage of about 100 V which can be used in a high-voltage work was obtained with Zener diodes. An idea for the PC programmable instruments of the old instruments was proposed.