Journal of the Japan Society of Applied Electromagnetics and Mechanics
Online ISSN : 2187-9257
Print ISSN : 0919-4452
ISSN-L : 0919-4452
Academic Papers
Measurement of Coil-to-coil Capacitance and High-frequency Patient Leakage Current in High-frequency Isolation Transformers for Transcutaneous Energy Transmission Systems
Investigation with Variations in Ferrite Core Number and Size
Tomoka OISHI, Kazuto SUGAI, Kenji SHIBA
Author information
JOURNAL FREE ACCESS

2026 Volume 34 Issue 2 Pages 128-134

Details
Abstract
High-frequency isolation transformers are used to suppress high-frequency patient leakage currents generated in medical equipment. This study analyzed the coil-to-coil capacitance and power transmission efficiency when varying the number (1 or 3) and size (φ25.3 mm or φ34.0 mm) of ferrite cores used in the high-frequency isolation transformer. Analysis results showed that using three φ25.3 mm cores (S3) yielded coil-to-coil capacitance C of 2.84 pF and power transmission efficiency η of 97.85 %. Furthermore, when using three φ34.0 mm cores (L3), C was 4.18 pF and η was 97.21 %, indicating that increasing the number of ferrite cores reduced C. When this high-frequency isolation transformer was inserted into the transcutaneous energy transmission system, for S3, the high-frequency patient leakage current Ip was 6.50 mA and the power transmission efficiency η' was 72.35 %. For L3, Ip was 6.61 mA and η' was 72.92 %, both falling below the regulation value of 10 mA.
Content from these authors
© 2026 The Japan Society of Applied Electromagnetics and Mechanics
Previous article Next article