Abstract
As acid-base balance should be taken into consideration for realizing a rapid and comfortable therapy, we have measured permeability of HCO3- and CH3COO- across membranes of dialyzers and found that PMMA membrane dialyzer, “B2-300” has significantly lower permeability for these two solutes than a cellulosic dialyzer with same surface area despite of its comparable permeability for urea and creatinine. It seems that such reduced permeability of “B2-300” prevents rapid loss of HCO3- into and inflow of CH3COO- from dialysate containing CH3COO- and that it thus contributes to a decrease in dialysis-related symptoms observed in the comparative clinical evaluations.