Comparison of hematology between Bio-Pump and Roller Pump
K. KOJIY. KUNIYOSHIF. IKEMURAK. IHAA. KUSABAM. SHIMABUKUROT. KAMIZATOY. TOYODA
Author information
JOURNAL FREE ACCESS

1986 Volume 15 Issue 2 Pages 545-548

Details
Abstract
This study was done to determine the effect on blood components during extracorporeal circulation between Bio-Pump and Roller-pump. During six hours bypass plasma free hemoglobin increased a little in control group (1.88mg/dl/h) and Bio-Pump group (1.85mg/dl/h). It increased significantly in Roller-Pump group (12.60mg/dl/h). Lactate dehydrogenase value showed same change of free hemoglobin.
Content from these authors
© The Japanese Society for Artificial Organs