Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
Manufacture of Simplified Bicarbonate Dialysis Unit Making Use of Acetate Dialysis Device
G. TAKAHASHIT. YOSIDAM. HIROSEM. MITACHIT. UTUMIYAA. OGAWAI. AMANO
Author information
JOURNAL FREE ACCESS

1986 Volume 15 Issue 3 Pages 1267-1270

Details
Abstract
We manufactured a Bicarbonate dialysis unit (Bc. unit) which is capable of performing also Bicarbonate dialysis, making use of the proportioning pump with which an existing Acetate dialysis device (Ac. device) has been equipped. Its external dimensions are 40(W) x 60(H) x 60(L)cm and capacity is 2.5/min. (1) When main stock solution containing no Bicarbonate is diluted in the Ac. device and poured into the quantitative tank within the Bc. unit, Bicarbonate solution is fed into the quantitative tank using a roller pump and agitated. Then, the concentration is checked forr supplying. (2) After this method has been used for dialysis, all functions properly operated and the electrolyte and pH of the dialysate were both stable.