The evaluation of PAN-17DX dialyzer mada of PAN membranso
J. MINAKUCHIM. MIZUGUCHIS. KAWASHIMA
Author information
JOURNAL FREE ACCESS

1991 Volume 20 Issue 1 Pages 82-84

Details
Abstract
The evaluation was made on PAN-DX which was improved by making PAN-CX2 thinner and gradient micro porous structure. Four hour hemodialysis was performed on 6 patients at QB 200ml/min and QD 500ml/min using PAN-17DX (PAN membrane, 1.7m2) Clearance Batas of low molecular weight substances were as follows, 172ml/min for BUN, 145ml/min for creatinine, and 135ml/min for IP. As for filtration characteristics, SC for β2MG and albumin were 0.51 and 0.001 respectively, whereas removal proportion of β2MG was 55.9% and its clearance amount into dialysate was 143.5mg, It was thought that PAN-17DX was high performance membrane with sharp cutoff characteristics.