Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
The Echo Doppler and Hematological Assessment of CarboMedics Prosthesis
—Comparison with Björk-Shiley Valve, Duromedics Valve, and St. Jude Medical Valve—
Kazuhiko SUZUKISousuke MIYAZAWAYoshimasa SAKAMOTOFumie SAITOHShigeru SUZUKIShigeki HORIKOSHITatstita ARAIHiromi KUROSAWA
Author information
JOURNAL FREE ACCESS

1992 Volume 21 Issue 2 Pages 601-605

Details
Abstract
We compared the hemodynamic performance of post mitral valor replacement among 4 mechanical valves. Recently, we use the CarboMedics prosthesis in mitral valve replacement. Between 1989-1990, 23 CarboMedics valves were implanted. Instead of this, we used Björk-Shilcy valves, St. Jade Medical valves, and Duromedics valves a short time ago. In evaluation of valve function, we measured peak flow velocity, pressure gradient, pressure half time, and mitral valve area by using Doppler echocardiography. Furthermore with regard to intravascular hemolysis, we evaluated the hematological effects after operation.
In conclusion,
1) The CarboMedics valve has a well hemodynamic function comparable to other 3 mechanical valves in an assessment from Doppler study. 2) With regard to hematological effects, lactate dehydrogenase, diret biliruhin, and indirect biliruhin are statistically lower in the CarboMedics valve than the Duromedics valve.
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback