Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
EVALUATION OF THREE SMALL SIZE ADULT TYPE HOLLOW FIBER MEMBRANE OXYGENATORS
K. INOUEH. AKAGIY. TAENAKAT. NAKATANIE. TATUMIM. MATUOM. KINOSHITAY. BABAT. MASUZAWAM. GOTOE. SASAKIH. ANAIM. SAKAKIK. ARAKIH. TAKANO
Author information
JOURNAL FREE ACCESS

1992 Volume 21 Issue 2 Pages 800-804

Details
Abstract
Gas exchange performance, heat exchange efficiency and pressure (trot) of new type small size membrane oxygenator ( UNIVOX: surface area 1.6 m2, priming volume 220 ml, MElkA EXCELLING a HPO-15RHF :1.5m2, 210 ml, MONOLYTH : 2.2 m2, 290 ml) were evaluated, and were compared with CAPIOX-II 54. Oxygen transfer performance and CO2 transfer performance of three oxygenator were excellent as like the CAPIOX-II 54. Three oxygenator have excellent heat exchange efficiency and low pressure drop. This study indicates that three new oxgenators are suitable for clinical application particularly for clear primed extracorporeal circulation. The control index of each oxygenator at 3 L/min of blood flow rate were 0.9-1.1 (UNIVOX), 1.0-1.4 (MERA EXCELUNG α HPO-15RHF), 0.9-1.2 (MONOLYTH), respectively.