人工臓器
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
吸着型人工腎とREDY SYSTEM
吉山 直樹中川 成之輔越川 昭三
著者情報
ジャーナル フリー

1974 年 3 巻 1 号 p. 33-38

詳細
抄録
Recent advances in artificial kidney design would suggest the possibility of small portable and wearable types. The most attractive one is a therapeutic application of substances capable of adsorbing solutes. which are pathophysiological significance in uremia.
Our programe had been undertaken in developing newer sorbent-based type dialysis, one of which is directly perf using over column of activated charcoal.
On the other hand, REDY SYSTEM (CCI Life System, Inc.) is a type of sorbent-based equipment regenerateting dialysate, and clinicaly evaluated in USA.
This unique dialysate supplying system is operated only 5.5L of dialysate.
This paper is a clinical report of this system in maintaining long-term hemodialysis patients.
As a result, we encountered no undesirable complications or hazardous accidents in 29 dialysis. Patients were able to keep good salt-water balance, body weight and acid-base balance. But in some instances, we came across hypopotassemia in using those poorbuild.