Expanded polytetra-fluoroethyleneの人工血管への応用
第3報静脈系における抗血栓性とporosityと表面構造
松本 博志木村 泰三布施 勝生山本 光伸三枝 正裕高松 俊昭深田 栄一
著者情報
ジャーナル フリー

1974 年 3 巻 5 号 p. 337-341

詳細
抄録
Our comparative studies of fluorinated materials which was used as the vascular prostheses for large caliber veins disclosed that surface properties and porosity highly corresponded with the thickness of the layer of thrombus in the inner surface. In the nonoporous prostheses (polytetrafluoroethylene-hexafluoropropylene, polytetrafluoroethylene-hexafluoropropylene-methylmetacryate.) initial platelet adhesion was seen from 24 hours to 120 hours after insertion into superior vena cava and inferior vena cava, and thrombus formation was seen in only one case during this period. In expanded poly-tetrafluoroethylene prostheses of 2-4 microns pore size, thin layer of thrombus which might be changed into the neointima was seen. On the other hand in ex-panded polytetrafluoroethylene prostheses of 7-10 microns pore size, thick layer of thrombus which could be enough to occlude the lumen was seen.