人工臓器
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
代謝系および内分泌系人工臓器用マイクロポンプの開発
壁井 信之菊地 真山田 明夫町山 悦子桜井 靖久
著者情報
ジャーナル フリー

1978 年 7 巻 1 号 p. 105-108

詳細
抄録

The roller type micropump driven by the pulse motor was developed as the insulin infusion pump for artificial cell. Flow rate of this pump is able to control by output signals of microcomputer. The volume displacement for one pulse was about 0.006μl. The pump output was increased in quite linear fashion from 0.003 to 0.1ml/min as increase the pulse number. The reproducibility was better than 2 % and pump flow rate was independent of load pressure.
As the next version of this type of pump, smaller one was also developed. Dimensions and weight of this pump are almost same as those of an usual wrist watch, so that it can be put on the patient's hand.

著者関連情報
© 一般社団法人 日本人工臓器学会
前の記事 次の記事
feedback
Top