人工臓器
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
左心補助心臓のUCGによる評価
仁田 新一香川 謙佐藤 尚加畑 治本田 剛彦本郷 忠敬毛利 平堀内 藤吾田中 元直目黒 泰一郎
著者情報
ジャーナル フリー

1978 年 7 巻 1 号 p. 70-73

詳細
抄録
An echocardiographic evaluation of right and left ventricular function during synchronous and asynchronous left heart device pumping was carried out in 20 mongrel dogs to establish the long term monitoring system as a non-invasive method.
The echo transducer was placed on the right ventricular epicardium directry or 1.5-2cm away from it, as the immersed method. The effects of LHB on heart rates, LVDd, LVDs, RVDd, RVDs and LVSV were measured and following results were obtained; 1) almost no changes of heart rates and RVD during sync. and async., 2) a significant decrease of LVDd and LVSV during both sinc. and async. with good correlation to the bypass ratio, 3) no signifcant changes of these determinants between async. and sync.
The results of this study demonstrate the usefulness of echocardiographic monitoring in the evaluation of a left heart assist device as a noninvasive, repeatable and reliable method.
著者関連情報
© 一般社団法人 日本人工臓器学会
前の記事 次の記事
feedback
Top