Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
DEVELOPMENT OF A CLOSED CIRCUIT OF HEART LUNG MACHINE WITH PULSATILE PUMPS
Naondo TAKIDOIwao FUJIMASAKou IMACHINorishige IWAHitoshi MIYAKEAkimasa KOUNOKazuhiko ATSUMI
Author information
JOURNAL FREE ACCESS

1978 Volume 7 Issue 5 Pages 689-692

Details
Abstract
A closed circuit of pulsatile flow pump-oxygenator named “Pulsatile Heart Lung Machine” (PHLM) was developed.
This system is consisted of two air driven pulsatile pumps, a membrane, a heat exchanger, a pump driving device, a halothane vapourizer, an electromagnetic flowmeterand a suction pump.
the caracteristics of this system are as follows;
1) pulsatile inflow from vein and outflow to artery.
2) relatively small priming volume.
3) good and easy controlability
4) easy to get sufficient perfusion flow during cardiac arrest.
5) compact machine in size.
This system will be able to be used in clincal long term extracorporeal circuration.
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback