体外循環に伴なう溶血の一評価法
酒井 清孝池田 博之本多 秀夫鈴木 満東 仲宣飯坂 修
著者情報
ジャーナル フリー

1979 年 8 巻 1 号 p. 59-62

詳細
抄録

Some in vitro tests of hemolysis using the experimental circuit were carried out in order to determine the hemolysis factors. It was investigated that occlusion of roller pump and hematcrit affected the hemolysis, and prevention of blood turbulent flow was useful in decreasing the hemolysis. Further, hemolysis model and rate equation were deviced, and by which above data were correlated. This hemolysis model and rate equation were found to be able to expain the hemolysis mechanism.