臨床化学シンポジウム
Online ISSN : 2187-4085
Print ISSN : 0386-3417
ISSN-L : 0386-3417
B-5. 高速液体クロマトグラフィーによるグアニジノ化合物の測定
細坪 秀夫古川 一郎有末 一隆甲田 一馬林 長蔵宮井 潔石田 泰夫
著者情報
ジャーナル フリー

1980 年 19 巻 p. 101-105

詳細
抄録
Guanidino compounds which are one of the uremic toxins are determinated by High Performance Liquid Chromatography (HPLC).
We investigate some conventions to perform with small volume of sample and within short time using HPLC. A length of column is shortened to 1/3 of conventional one, that is 5cm of column length and the analysis of peaks from taurociamine to methylguanidine is shortened to 1/5 (26 min) of conventional one.
A volume of sample used only 1000μl for an analysis. The lineality of standard compounds and precision are also good as well as results with conventional method, so that this method could be used for clinical test.
著者関連情報
© 日本臨床化学会
前の記事 次の記事
feedback
Top