多変量分析法による個別検体の精度管理法
臼井 敏明
著者情報
ジャーナル フリー

1977 年 5 巻 4 号 p. 391-396

詳細
抄録
A method of quality control for the individual sample is presented by use of amultivariate statistical method. Multiphasic clinical chemistry data are processed by principal component analysis, and components with small variation are taken as indicators of the quality control. Abnormal values of these principal components are indicative of either unusual pathological states or mistakes of laboratory technics. The quality control with this method corresponds well to the data-check by the clinical pathologist.