臨床化学
Online ISSN : 2187-4077
Print ISSN : 0370-5633
ISSN-L : 0370-5633
Amberlite XAD-2を用いたエストロゲン測定用キットの問題点
山口 賀久有末 一隆林 長蔵
著者情報
ジャーナル フリー

1980 年 9 巻 3 号 p. 355-357

詳細
抄録
We evaluate E3 kit which is commercially available, and have found some problems. Effect of protein on adsorption of estrogen to Amberlite XAD-2 is observed and the effect is increased by increased concentration of protein.
Abnormal absorption curve is also observed with samples from pregnant woman given some antibiotics. The interference of these drugs is corrected by three points wavelength measurement, but the effects of protein could not be avoided.
著者関連情報
© 日本臨床化学会
前の記事 次の記事
feedback