熱分析を用いた医薬品候補化合物の迅速な安定性予測法
後藤 晃範 森 浩俊
著者情報
ジャーナル フリー

2011 年 38 巻 4 号 p. 112-115

詳細
抄録
For thermoanalysis-based stability prediction of a drug candidate, the highest temperature applicable so far to humidity-controlling of gas flow into a heat furnace has been limited to about 80℃. We solved this limitation by placing a water-containing bottle between gas-flow tubing segments and then flowing a humidity-saturated nitrogen gas at room temperature into the heat furnace to decompose a sample. This allowed furnace-heating at 150℃. In this study, the predicted decomposition rate was compatible with the actual decomposition rate, but it would be necessary to validate this method in future.