レゼクトスコープ施行中にカッティングループの破損を生じた一例
小迫 優子桜井 友義佐野 陽子佐々木 宏輔大村 伸一郎原 澄子
著者情報
ジャーナル フリー

2010 年 26 巻 2 号 p. 444-447

詳細
抄録

We experienced The breakage of resectoscopic cutting loop tip while we were performing the transcervical resection . And the broken pieces were all retrieved without causing any trouble. It is very rare to break a loop tip, but a cutting loop should be maintained in a good condition at anytime. And it is mandatory to check it whether if there is any crack or worn-out before starting the operation to prevent this kind of complication. Also, we learned the importance of knowing how to immediately manage sudden break of the resectoscopic cutting loop.