腹腔鏡下手術における分流熱傷の新しい機序を解明した1例
-絶縁体である回収袋内でモノポーラーの通電を生じたことによる-
近藤 壯塩野入 規塩沢 功
著者情報
ジャーナル フリー

2011 年 27 巻 2 号 p. 412-415

詳細
抄録
  Monopolar current is essential in laparoscopic surgery. However complications of monopolar electrosurgery include thermal injury due to wrong contact, capacity coupled current, and shunting . This report deals with a new mechanism of shunting in laparoscopic surgery. The specimen bag is the insulator. In this case, the monopolar current cut tumor in the collecting bag. Typically this can't happen. Electricity should have flowed to the surface of the abdomen. When the bag had full contact with organs it did not cause shunting burns. If there was point contact, shunting burns occurred. We propose that the causes of shunting burns are point-contact and the water in the abdominal cavity. Electrothermal burns are reported more often in laparoscopic surgery than in laparotomy. This report aims to reduce these injuries through understanding of the characteristics of the electric scalpel and water management in the abdominal cavity.