日本リウマチ・関節外科学会雑誌
Online ISSN : 1884-9059
Print ISSN : 0287-3214
ISSN-L : 0287-3214
高線量γ線照射がポリエチレンの摺動面ならびに摩耗粉におよぼす影響
山本 謙吾, 宍戸 孝明, 正岡 利紀, 伊藤 康二, 今給黎 篤弘, I.C. Clarke, 東海林 宏, 川那辺 圭一, 大西 啓靖
著者情報
ジャーナル フリー

2002 年 21 巻 3-4 号 p. 225-232

詳細
抄録
PE cups with 0, 2.5, 50, 100 and 150 Mrad radiation treatments were run in a hip simulator, and we examined the cup surfaces for micro-wear phenomena and analysed the corresponding wear debris. In general, the size and frequency of the PE surface fibrils and the size of the retrieved PE debris decreased with increasing radiation dose. The PE fibrillar sizes and geometrical shape-factors on the cup surfaces correlated well with radiation dose. However, the trends for debris size and shape-factors as related to radiation dose were weak. Thus, the morphology of the PE-fibrils on the cup surfaces was more sensitive to variations in radiation dose than the actual wear debris produced in simulator tests.
著者関連情報
© 日本リウマチ・関節外科学会(2006年~:2005年以前は投稿規程に著作権に関する記載なし)
前の記事 次の記事
feedback
Top