グリッドアクティブステレオを用いた三次元消化器内視鏡におけるキャリブレーション手法
青木 広宙古川 亮西谷 維心青山 正人日浦 慎作小南 陽子松尾 泰治吉田 成人田中 信治佐川 立昌川崎 洋
著者情報
ジャーナル フリー

2014 年 52 巻 2 号 p. 97-107

詳細
抄録
Calibration for calculating the external parameter of camera-projector system is important, to make high-precision 3-D measurement by 3-D endoscope using the grid active stereo. In this paper, we propose a new calibration method with a sphere calibration object in a geometrically flexible camera-projector system, and we examine the validity of the proposed method. As the result of comparing using conventional calibration with a plane calibration object, it is clarified that the proposed method resolves the scaling imprecision, that is a problem of the conventional method. Thus, the proposed method has higher potential of measurement accuracy. In addition, it is the advantage of the proposed method that precision estimation of calibration parameters with a low number of input images is possible.