生体医工学
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
一般演題(口演)
直接給電FES用体外給電システムの軽量化モデル設計に関する検討
菅 一将佐藤 忠邦田倉 哲也佐藤 文博松木 英敏
著者情報
ジャーナル フリー

2014 年 52 巻 Supplement 号 p. O-515-O-516

詳細
抄録
FES is the therapy for rehabilitate lost movement function to apply electrical stimulation to paralyzed extremity. We adapted implanted direct feeding method for FES. Direct feeding method is the stimulus method that transmits the energy for driving implants and communicates bi-directionally by electromagnetic induction with a mounted coil and implant coils contactlessly.However, we guess mounted system is too heavy for paralyzed patients. So we must reduce the weight.We proposed the new mounted coil with plural repeaters, the circuit composed of a capacitor and a coil with series connection. We designed mounted coil that can drive implanted coil by controlling a current in each repeater and magnetic field generated from a mounted coil.Consequently, we realized the 18turns lighter mounted exciting coil than the traditional coil and confirmed this coil can generate necessary magnetic field, 0.35mT, to implanted coils.
著者関連情報
© 2014 社団法人日本生体医工学会
前の記事 次の記事
feedback
Top