Transactions of Japanese Society for Medical and Biological Engineering
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
Organized Session
Development of Next-Generation Cascade Blood Pump for Total Artificial Heart
Kazuyoshi FukunagaMasaya OnukiItsuro SaitoTakashi IsoyamaYusuke AbeTomoyuki YambeAkio FunakuboYasuhiro Fukui
Author information
JOURNAL FREE ACCESS

2014 Volume 52 Issue Supplement Pages OS-3-OS-4

Details
Abstract
We have aimed to apply a cascade pump, which has a high ability to generate a pulsatile flow, to a total artificial heart. In a cascade pump, the pressure increases along the flow passage, and therefore the eccentric force occurs, which moves the rotating impeller from the outlet to the inlet. To reduce the force, we have developed a new cascade pump in which one flow from the inlet separates into four. This special structure enables the pressure distributions within the pump to be symmetric with respect to the X, Y, and Z axes. The pump performance test with bovine blood demonstrated 10 L/min against 170 mmHg at 3,200 rpm. Furthermore, we developed a system for measuring the eccentric force on the impeller. The eccentric force was 0.2 N on condition that the pump provided 9.3 L/min against 100 mmHg at 2,500 rpm.