モデル誤差を用いた補助人工心臓運用時の拍動推定と同期制御
幡谷 原太田中 明吉澤 誠白石 泰之三浦 英和山家 智之
著者情報
ジャーナル フリー

2015 年 53 巻 Supplement 号 p. S196_03

詳細
抄録
It is reported that a synchronized control of rotary left ventricular assist devices has some advantages in comparison with constant rotational speed control. The most common way to synchronize to heart beat is to use ECG signal because R-wave represents the beginning of systole and is easily detected. However, it is difficult to detect the onset of diastole. The aim of this study is to develop the algorithm to synchronize to heart cycle without external sensors such as ECG.In proposed method, a heartbeat was regarded as disturbance to pump rotational speed. Thus the disturbance was estimated as the difference between measured and estimated rotational speed. An ARX model which input was the reference of pump speed and current consumption was adopted to estimate the instantaneous rotational speed. The proposed method was tested in both a mock circulatory system and animal experiments. The results indicated that the system can output the heart cycle signal and detect not only the onset of systole but that of diastole. Furthermore, co-pulse and counter-pulse mode control can be demonstrated by using the detected timing. However there was detection delay at start of systole.