PET動態画像の4次元再構成における基底関数の種類による画質の変化
山田 純也本谷 秀堅木村 裕一坂田 宗之
著者情報
ジャーナル フリー

2015 年 53 巻 Supplement 号 p. S214_02

詳細
抄録
The authors studied the quality of dynamic PET images reconstructed from not a sinogram measured at each time but a set of sinograms measured during the observation. Representing a time activity curve observed at each voxel using set of basis functions, the authors reconstructed the images by estimating the coefficients for linear sum of the basis functions that maximize the likelihood. The likelihood is defined under the assumption that the measurement obey the Poisson distribution and maximized by means of an EM algorithm. In this presentation, the authors report that you can improve the image quality by simulnaneously reconstructing dynamic PET images from a set of all sinogram data.