レーザ超音波法を用いた完全非接触光音響計測による鶏ささみ肉内の模擬血管イメージング
山本 摂三浦 崇広千星 淳落合 誠
著者情報
ジャーナル フリー

2015 年 53 巻 Supplement 号 p. S218_03

詳細
抄録
In blood vessel imaging using photoacoustic effect, ultrasonic signals which are induced by a laser are commonly detected by conventional contacting transducers. Replacing contacting transducers with laser-ultarsonic measurement using laser interferometer, complete non-contact biological tissue imaging would be realized. Non-contact measurement technique would enhance the applicability of ultrasonic measurement. Authors have developed non-contacting photoacoustic imaging technique using Fabry-Perot interferometer. As a generation laser and a detection laser, Nd:YAG laser (2nd harmonic wavelength: 532 nm) and Nd:YAG laser (fundamental wavelength: 1064 nm) were adapted respectively. To improve the sensitivity of detection, 3 dimensional frequency-domain synthetic aperture focusing technique was developed. For the verification, chicken tender and small-bore tubes with an inner diameter 0.1 mm filling with red ink solution were used as a biological tissue and artificial blood vessels respectively. Using non-contact measurement technique, artificial blood vessels which installed under chicken tender with 6 mm thickness were clearly observed.