臨床応用に向けた内シャント光イメージングの基礎的検討 —画像差分による血管透視像の明瞭化—
神山 英昇北間 正崇清水 久恵山下 政司横山 徹小島 洋一郎清水 孝一
著者情報
ジャーナル フリー

2017 年 55Annual 巻 Proc 号 p. 593-594

詳細
抄録

We have proposed a noninvasive technique to observe an arteriovenous fistula using an optical transillumination image of a blood vessel. To improve the accuracy of inner-diameter measurement of the vessel, we examined the effectiveness of the imagesubtraction technique using multi-wavelength light sources. In the experiment, we fabricated a model-phantom to simulate a human forearm. It contained the tube filled with replaceable absorber liquid which simulated the blood in different wavelengths of illuminating light. The transillumination images of the blood column were recorded with the liquid of different absorption coefficients. The imagesubtraction technique was applied and the images were analyzed. With the subtraction, the sharpness of the column image was improved by 55%, and the error in the diameter measurement was reduced from 28% to 4%. This result verified the effectiveness of the image-subtraction to promote the clinical application of the proposed technique for the controlled care of the arteriovenous fistula.

著者関連情報
© 2017 社団法人日本生体医工学会
前の記事 次の記事