バイオエンジニアリング講演会講演論文集
Online ISSN : 2424-2829
セッションID: 2B22
会議情報
2B22 ヒト血液を用いた持続的血液濾過器のin vitro血栓性評価試験法の開発
平田 麻由紀岩崎 清隆松橋 祐輝高橋 東山本 祥宜鮫島 啓青山 祐介梅津 光生
著者情報
会議録・要旨集 認証あり

詳細
抄録
Hemofilters are medical devices for assisting deteriorated renal functions, which purify the blood by removing pathogen and wastes. For continuous hemofilters, antithronbogenecity is one of the most important requirements. The aim of this study is to investigate the key parameters for the establishment of a reliable in vitro thrombogenecity testing methodology for continuous hemofilters using human blood from healthy volunteers. This study used adjusted activated clotting time of human blood drawn from healthy volunteers to 150-200 sec using heparin. The blood was circulated in the air-contactless circuit with flow and pressure in a clinical situation was reproduces. The tests were terminated when the inlet pressure elevated to 150 mmHg from the initiation pressure of 70 mmHg at the inlet. The average circulation duration was 6.5±1.7 hours(n=4). The circulation duration using human blood from healthy volunteers were shorter than using porcine blood which was previously reported. Hematocrit of porcine blood was lower than the human blood from healthy volunteers, whereas the hematocrit of porcine blood was comparable to dialysis patients. These data indicated the importance of hematocrit of blood in the assessment of the thrombogenecity of continuous hemofilters.
著者関連情報
© 2016 一般社団法人 日本機械学会
前の記事 次の記事
feedback
Top