0329 狭い隙間を有する一軸偏心スクリュー血液ポンプに関する基礎的研究(OS3 流体機械の研究開発とそれに関連した複雑流動現象,オーガナイズドセッション)
重光 翔大倉 成夫堀江 昌朗
著者情報
会議録・要旨集 フリー

詳細
抄録
In our current research, the efficiency of a centrifugal blood pump for infants is about 30% or less at 2.0L/min. It is important to improve efficiency and to prevent blood trauma. In general, a positive displacement pump is suitable for the low specific speed. In this research, a single screw pump which has narrow gap between a rotor and a stator was designed to prevent blood trauma. The fundamental performance such as pump characteristic, leakage characteristic, and pump efficiency of this single screw blood pump was investigated.
著者関連情報
© 2012 一般社団法人 日本機械学会