流体工学部門講演会講演論文集
Online ISSN : 2424-2896
セッションID: OS13-1
会議情報

循環流路内における血栓形成の電気的計測およびシミュレーションによる評価
*上村 拓生小石 まどか李 建平Kiet Anh TRAN伊藤 恭介武居 昌宏
著者情報
会議録・要旨集 フリー

詳細
抄録

Connector gap, a gap between connector and tube, is where very easily to occur thrombus formation in extracorporeal blood circulating system. The aim of this study is to develop a system to detect thrombus formed at the connector gap in an extracorporeal blood circulating system and evaluate it. In the experiment, the blood circulating system incorporating the connector electrode was filled with fresh porcine blood, and calcium chloride was added to promote thrombus formation. Also, the experimental condition was modeled by simulation, the frequency response was obtained, and it was evaluated compared with the experimental results. In both the experiment and the simulation, the permittivity greatly increased at low frequencies, and the rate of increase of the permittivity showed the minimum value at about 10 MHz because of thrombus formation. The feasibility of the connector electrode and the usefulness of the simulation were shown.

著者関連情報
© 2018 一般社団法人 日本機械学会