ICONE15-10071 CALIBRATION OF A PROPORTIONAL COUNTER FOR DETERMINING BETA EMITTERS
Marko StrokUrska RepincBorut Smodis
著者情報
会議録・要旨集 フリー

詳細
抄録
Calibration of recently installed proportional counter at the Hot Cells Facility of the Jozef Stefan Institute was performed. Instrument was calibrated for determination of total beta activity, Sr-90 and Pb-210. Detection efficiencies for K-40, Sr-90, Y-90, Pb-210 and Bi-210 were determined, allowing for more accurate determination of the particular nuclide as a single K-40 efficiency. In addition, self-absorption curves for different surface densities for the nuclides mentioned were derived. Two empirical equations for faster and more accurate determination of Sr-90 and Pb-210 were derived. These two equations consider differences in surface density and in-growth of Y-90 and Bi-210 respectively. The detection efficiencies obtained ranged from 10 to 52 %, depending on the nuclide, surface density and chemical compositions of the salts used or precipitates obtained following radiochemical separation in the experiment. As an example, total beta activities in five samples of mineral waters according to the relevant ISO standard were determined. All procedures and formulae developed include calculation of minimal detectable activities and uncertainty budgets for the determinations concerned.