IIP情報・知能・精密機器部門講演会講演論文集
Online ISSN : 2424-3140
セッションID: F-04
会議情報

フレキシブルコイルを用いた人工心臓用磁界共振結合経皮電力伝送システムのファントムによる評価
*長戸 誠良土方 亘進士 忠彦
著者情報
会議録・要旨集 フリー

詳細
抄録
A conventional power supply method using a drive line for a ventricular assist device causes infection diseases, which decreases patient’s quality of life and survival rate. A possible solution is to use a transcutaneous energy transmission system utilizing electromagnetic induction. However, it is necessary for the coil to be located on the body surface accurately because of short transmission distance. In order to avoid infection diseases as well as increase the transmission distance, we proposed the new system utilizing magnetic resonant coupling that embedded flexible coils into clothes to transfer energy. In this paper, we evaluated the influence of the electric characteristics of the human body on the transmission efficiency by using both phantom and pork. The measured transmission efficiency with a gap of 30 mm was 67 % in phantom and 70 % in pork while that in air was 80 %.
著者関連情報
© 2017 一般社団法人 日本機械学会