主催: 一般社団法人 日本機械学会
会議名: 日本機械学会 関東支部第28期総会・講演会
開催日: 2022/03/14 - 2022/03/15
A spectroscope is used to measure emission intensity for each wavelength in Laser-Induced Breakdown Spectroscopy (LIBS), therefore its measurement system is large. However Laser-Induced Plasma Spectroscopy (LIPS) can measure distributions of concentration and temperature of the atoms and molecules by taking the plasma images only through optical filters. In this study, emission spectra from H, O atoms and CN, C2 molecules were calculated and these concentration and temperature measurements by LIPS were simulated from the calculated spectra.