年次大会講演論文集
Online ISSN : 2433-1325
セッションID: 601
会議情報
601 ヒートパイプを用いた体外冷却型異常放電抑制装置 : 基本性能の検討(J04-1 手術・治療,J04 ライフサポート)
内山 城司中野 公彦斉藤 俊安田 利貴佐伯 壮一田中 信宏藤井 正美
著者情報
会議録・要旨集 フリー

詳細
抄録
The cooling of brain tissue has been recognized to effectively suppress epileptiform discharges (EDs) and local brain cooling using a thermoelectric device has recently gained much attention since it has a potential to be a viable alternative for a surgical resection of epileptogenic foci. However, when considering implantable systems, several problems such as the electric power supply or the arrangement of the whole system in the body have to be resolved. Then, the authors devised a new cooling device with a heat pipe attached to the thermo-electric chip and also examined the performance of this device through the experiments using a phantom and animal studies. The results indicate the proposed device has enough cooling performance to suppress EDs.
著者関連情報
© 2006 一般社団法人日本機械学会