年次大会講演論文集
Online ISSN : 2433-1325
セッションID: F-0722
会議情報
F-0722 回転型剪断負荷装置を用いた力学特性模擬血液の溶血特性(J22-3 生体関連の熱・流体工学(3))(J22 生体関連の熱・流体工学)
丸山 修山根 隆志西田 正浩筒井 達夫軸屋 智昭増澤 徹
著者情報
会議録・要旨集 フリー

詳細
抄録
We are developing microcapsule suspension, which is expected to be used for absolute hemolytic evaluation instead of animal blood. In this report, we describe fractural characteristics of microcapsule suspension under fluid dynamical shear stress, which is generated by rotational shear stressor developed in our laboratory. We tested two types of microcapsule suspensions with 100 μm diameters. One has polyurethane membrane, and the other has melamine-resin membrane. Fractural property of the polyurethane suspension was similar to that of bovine blood, whereas the melamine-resin suspenstion showed different result. Therefore the polyurethane suspension will be applicable to absolute hemolytic evaluation of centrifugal blood pumps.
著者関連情報
© 2001 一般社団法人日本機械学会