骨足場へのEr:YAGレーザ照射時の温度測定
天野 裕晴古本 達明橋本 洋平小谷野 智広細川 晃
著者情報
会議録・要旨集 フリー

詳細
抄録

It is expected that the removal operation of the calcified heart valve will be simpler if we use lasers in substitution for drills. In this study, surface temperature of bone scaffold including hydroxyapatite during pulsed Er:YAG laser irradiation was measured using infrared radiation pyrometer as fundamental experiment. As a result, the surface temperature during laser irradiation became more than 60°C that is a glass transition point of the polylactic acid. Then, polylactic acid softened. The temperature is associated with the removal volume. Additionally, we can suppress the heat storage to the sample by regulation the pulse interval of the laser.