マイクロ・ナノ工学シンポジウム
Online ISSN : 2432-9495
セッションID: 30am2-PN-42
会議情報
30am2-PN-42 埋込み型人工腎臓におけるマイクロ流路のバイオファウリング
大田 能士藤 直也森田 伸也菅野 義彦三木 則尚
著者情報
会議録・要旨集 フリー

詳細
抄録
Implantable artificial kidney can drastically improve the quality of life of the renal disease patients. In previous study, our group developed a multi-layered micro dialysis device which is composed of micro stainless steel channels and nano-porous PES membranes. In this study, we investigate the relation between the shape of channels and biofouling to provide long-term performance. And we evaluate the aged deterioration of the channel's surface for 1 week using human whole blood in reflux system. The result of this experiment indicated the relation between radius of curvature and biofouling. To optimize the shape of channels is crucial in order to implant this device in medium-sized animal in near future.
著者関連情報
© 2015 一般社団法人 日本機械学会
前の記事 次の記事
feedback
Top