胎児皮膚へのパッチ固定に向けたピンの構造変化による穿刺特性の評価
*中島 広夢土肥 健純桑名 健太
著者情報
会議録・要旨集 認証あり

詳細
抄録

In this study, we propose a structural changeable pin as a fixation mechanism for a treatment patch of Myelomeningocele, which is a fetal stage disease. The mechanism that is easy to insert and hard to extract can be realized by changing the pin structure after inserting into the fetal body. In this paper, a large scale model was prototyped and the insertion and the extraction forces were compared with those of a normal pin. The insertion forces of the normal pin and the prototyped pin were 1.7 N and 3.0 N, respectively. The extraction forces of the normal pin and the structural changed pin were 1.5 N and 3.0 N, respectively. These results indicate that the mechanism that is hard to extract can be acquired by using the proposed method.

著者関連情報
© 2018 一般社団法人 日本機械学会