ロボティクス・メカトロニクス講演会講演概要集
Online ISSN : 2424-3124
セッションID: 1A2-J12
会議情報
1A2-J12 A Microsurgery Robotic System with Wrist Action for Fingertip Replantation Surgery
KyuJin JOJongwon LEEYoungjin PARKWan Kyun CHUNG
著者情報
会議録・要旨集 フリー

詳細
抄録
In this paper, we developed a tele-operation robotic system for fingertip replantation surgery, which is called MiRS (Microsurgery Robot System). There are some difficulties on surgery process, because of tiny blood vessels and nerves of the finger. To cover the problems of surgery, the system, which is called MiRS (Microsurgery Robot System), must have two rightful prerequisites: the accuracy and comfortableness. For accuracy, which guarantees safety also, we developed a master-slave surgical system. Safety problem contains the reduction of hand-tremor on surgery, so "motion scaling" can be realized by master-slave system. Besides, motion scaling can improve the accuracy itself, either. Lastly the master-slave system with scaling motion improves comfortableness either. Most importantly, the wrist motion is added in end-effector for our system. That motion can helps make more natural and smaller movement on surgery. The process of surgery can be more safe and accurate by making small movement. We developed slave system now, and experimental results are included to verify the performance of the robot and advantages by using wrist on the end-effector.
著者関連情報
© 2009 一般社団法人 日本機械学会
前の記事 次の記事
feedback
Top