2P1-A07 The development of the area-variable compression cone for X-ray diagnosis of the stomach(Medical Robotics and Mechatronics)
Shintaro KAMEYAMAKoichi SUZUMORIShuichi WAKIMOTOHisao OKAKoichi SHIBUYAKazuyuki SUMIYayoi YONEZAWA
Author information
CONFERENCE PROCEEDINGS FREE ACCESS

Details
Abstract
Physical pressure with the compression cone is effective to get clear tripe and fold image in the X-ray diagnosis of the gastric region. However, moving the compression cone during pressing body would be dangerous actions such as bone fractures. In this study, we have developed the area-variable compression cone mechanism using the expansion of the balloon. The fabricated compression cone could pressure safely because a balloon has high flexibility. We confirm this effectiveness by pressuring the phantom of the stomach.