3P2-I08 長期にわたるマイクロ透析装置の使用評価(ナノ・マイクロ流体システム)
真田 一平伊藤 昂Gunawan Setia PrihandanaGu Ye登坂 万結鈴木 哲也菅野 義彦三木 則尚
著者情報
会議録・要旨集 フリー

詳細
抄録
We have ever developed a micro dialyzer consisting of micro channels and nano-porous polyethersulfone (PES) membrane for dialysis membrane. PES membrane is broadly used as a membrane for nanofiltetration such as fractionation, purification, and dialysis. However in the long term, PES membrane is apt to absorb and make the protein layer, which produces the platelet adhesion when its surface is contacted to the blood, then the performance of our device drastically get worse. Therefore in this work, we modified the PES membrane by coating Fluorinated Diamond Like Carbon (F-DLC) film. We investigated anticoagulants of the F-DLC/PES membrane using a microfluidic device and experimentally showed that the amount of cells absorbed to the surface substantially reduced in F-DLC/PES membrane. We also conducted long term use evaluation of F-DLC/PES membrane for 9 days and the performance of normal PES membrane was seemed to drop off while it was steady in F-DLC/PES membrane.