高速応答水素炎イオン化検出器用サンプリングチューブ内の変動濃度場の数値解析
*保浦 知也清瀬 大士服部 博文田川 正人
著者情報
会議録・要旨集 認証あり

詳細
抄録

To measure concentration fluctuation accurately by a fast-response flame ionization detector (FFID), the frequency response function should be known, and measured signals need to be compensated appropriately. In the present study, we numerically analyzed the convection and diffusion equations of concentration developing in a laminar pipe flow. Then, we have obtained the response functions of the cross-sectional mean concentration at several downstream locations by varying the frequency of concentration fluctuation at the inlet of the pipe. It is found that the theoretical and numerical results are in good agreement. Based on the response function obtained, we may compensate the attenuated concentration signals rationally.