臨床応用後回収されたインプラント体―骨界面の微細構造学的研究
中川 寛一市之川 浩大山 賢司木村 世紀粟崎 洋有泉 祐吾淺井 康宏
著者情報
ジャーナル フリー

1994 年 7 巻 1 号 p. 77-83

詳細
抄録
HA-coated Dental Implant (Kyoceya, Japan) were extracted from mandibular bone for fracture of the upper structure after 44 months, including a loading period of 40 months. And it was prepared using a cutting-grinding system and studied with transmission electron microscopy. The HA-coated Dental Implant showed an intimate bone contact at the ceramic surface under a loaded condition in human that appeared to achieve biointegration at the ultrastructural level. However, detachment of HA crystals from the coated surface is evident. Crystals are situated inside the newly formed bone surrounding the implant and the cytoplasm of phagocytic cells.
著者関連情報
© 1994 公益社団法人日本口腔インプラント学会